  Level Meter  Distortion Meter
ÚÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³--ÚÄÄÄÄÄÄÄÄÄÄÄ¿ÚÄÄÄÄÄÄÄÄÄÄÄ¿--³   I have been working on one of these
³  ³ú          ³³ú          ³  ³   old AF automatic distortion meters.
³  ³ '.        ³³ '.        ³  ³   It is quite different to the usual
³--ÀÄÄÄÄÄÄÄÄÄÄÄÙÀÄÄÄÄÄÄÄÄÄÄÄÙ--³   fiddly variable Wien Bridge deep
³                              ³   notch type. So it is ideally suited
³        0                     ³   for setting up tape machines, where
³     Function     Distort     ³   the playback level and frequency will
³     [][][][]    [][][][]     ³   never be that steady.
³ O                          o ³
³       [][][][][][][][]       ³
³ O                          í ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÙ
 In        Level Range     Mains

Level Ranges:- 30V, 10V, 3V, 1V, 300mV, 100mV, 30mV, 10mV.
            :-  30,  20, 10,  0,   -10,   -20,  -30,  -40 dBV

Response    :-    10Hz - 250kHz @ -3dB

Distortion Filters 400Hz, 1kHz & External.

Distortion Ranges: 30%, 10%, 3%, 1% min to 0.2%
            & dBs -10 -20 -40 -40  min to -55dB

Harmonic Response: 1.5x Filter to 130kHz @ -3dB

Mains 230V. odd 3 pin 1cm square socket.

The automatic refers to the distortion measurement does not need the zero level
reference level to be set first. (But is must be reading 30-100% on the level
meter scale.) This is done not by a very low distortion AGC system as you might
think, but a cleaver meter movement with 2 coils and no return springs. One
coil is fed with rectified DC from the input level detector 1 and acts as the
return spring and the other is fed from an 2nd detector from a high pass filter
that removes all of the fundamental, so it just the harmonics. The resultant
meter deflection is the distortion fraction.

In ÚÄÄÄÄÄ¿ ÚÄÄÄÄÄ¿    ACÚÄÄÄÄÄ¿              ÚÄÄÄÄÄÄÄÄÄÄÄ¿
oÄÄ´LevelÃÄ´InputÃÄÂÄÄÄÄ´LevelÃÄÄÄÄÄÄÄÄÄÄÄÄÄÄ´LEVEL METERÃÄÄÄÄÄ¿
   ³Atten³ ³ Amp ³ ³    ³Det 1³ DC           ÀÄÄÄÄÄÄÄÄÄÄÄÙ     ³
   ÀÄÄÄÄÄÙ ÀÄÄÄÄÄÙ ³    ³     ÃÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ³
                   ³    ÀÄÄÄÄÄÙ                         ÚÄÄÄÄÁÄÁÄÄÄÄÄ¿
                   ³   ÚÄÄÄÄÄÄÄÄÄ¿                      ³ DISTORTION ³
                   ³ oÄ´400Hz HPFÃÄo  ÚÄÄÄÄÄ¿  ÚÄÄÄÄÄ¿  ³2 COIL METER³
                   ³   ÃÄÄÄÄÄÄÄÄÄ´    ³ Set ³  ³Level³  ÀÄÄÄÄÂÄÂÄÄÄÄÄÙ
                   À>oÄ´ 1kHz HPFÃÄo<Ä´Gain ÃÄÄ´Det 2ÃÄÄÄÄÄÄÄÙ ³
                       ÀÄÄÄÄÄÄÄÄÄÙ    ³ Amp ³  ³     ÃÄÄÄÄÄÄÄÄÄÙ
                     oÄ External ÄÄo  ÀÂÂÂÂÂÙ  ÀÄÄÄÄÄÙ DC
                                        ^ÄÄÙDist Range

It just uses simple RC transistor Amps and bridge diodes for the electronics.
The notches are sealed LC high Q units. (M derived?)

There are only 3 presets:- Level gain, Distortion gain, & input amp Bias.

DISTORTION MEASURING PRINCIPLES
                                      f1  f2
     ³  f1                     0dBÄ´   |  ÚÁÄÄÄÄÄÄ     ³  f1
Level³   ³  f2                     ³   |  ³ High       ³      f2
     ³   ³   ³  f3                 ³   |  ³ Pass       ³       ³  f3
     ³   ³   ³   ³  f4  f5   -60dB_³___|_/  Filter     ³       ³   ³  f4  f5
     ÀÄÄÄÁÄÄÄÁÄÄÄÁÄÄÄÁÄÄÄÁ         ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄ     ÀÄÄÄÄÄÄÄÁÄÄÄÁÄÄÄÁÄÄÄÁ
       Distorted Signal                 Filter            Filtered Signal

The high pass filter must completely remove the fundamental and all LF hum etc.
but leave all the harmonics intact. The detectors (especially the distortion
one) should be a true RMS type, as the waveform will generally be very complex
and nothing like a sine wave. The ratio of f1 to all the harmonics f2+++ is the
distortion factor.

Calibration is done by adding in a 2nd tone (e.g. 3-10x freq) via high value Rs
at say 10% of the level to indicate 10% distortion. The distortion meter
bandwidth can also be checked with this circuit.

1V 1kHz Äo-oÄÄÄ10KÄÄÂÄÄÄÂÄÄÄÄÄÄ¿
 Osc 1   o          ³   ³      ³
        _³_         ³   ³   Distortion
                    ³  DVM    Meter
100mV               ³  _³_    _³_
2-20kHz Äo-oÄÄÄ10kÄÄÙ
 Osc 2   o
        _³_

To check the HPF filter is working OK a very pure sine wave is needed, so an LC
tuned circuit after a signal generator is used. But I did find that > 0.5V @
resonance on the meter did make the inductor distortion rise > 0.01% !

  Osc         88mH      500mV
50mV 1kHz ÄÄÄ((((()ÄÄÂÄÄÄÄÄ¿ Hi Z
 low Z               ³  Distortion
    ³        0.28uF ===   Meter
   _³________________³_____³

MODERN ANSWER
Of course with a good quality PC sound card and a Furrier analysis program that
can add up the harmonics distortion measurement should be easy.  


See my bul "Marconi 2331A Distortion Factor" & "Simple 1kHz AF osc".


Why don't U send an interesting bul?

73 de John, G8MNY @ GB7CIP