OpenBCM V2.0.3 (Linux)

Packet Radio Mailbox

IZ3LSV

[San Dona' di P. JN]

 Login: GUEST





  
G8MNY  > TECH     21.08.08 20:04l 87 Lines 2378 Bytes #999 (0) @ WW
BID : 304_YB8EW
Read: GUEST
Subj: (8 Bit ASCII graphics use code pag
Path: IZ3LSV<IK3GET<IW2OHX<OE7XLR<OE9XPI<DB0CZ<DB0GE<DB0LJ<DB0OVN<DB0IUZ<
      DB0FBB<DB0ERF<DB0MGB<DB0TGM<DB0BLO<DK0BLN<DB0TUD<SR6BBC<SR9ZAA<SP7MGD<
      VK7AX<VE2RXY<KD4GCA<N2BQF<N9PMO<K9BBS<VE3UIL<VK2DOT<XE1FH<LU1FES<
      PY2DML<PY1AYH<CT1EJC<ED1ZAC<I0TVL<SV1CMG<ZL2BAU<GB7ESX<GB7BED<GB7IMK<
      GB7YFS<VK4TRS<VE2PKT<VE1DRG<I6QPL<N8DA<K0KCC<AA5JJ<VK6HGR<ON0AR<DB0RES<
      DK0WUE<7M3TJZ<JE7YGF<JA3PYC<JI3ZAG<YB4AC<YB8EW
Sent: 170316/0119Z @:YB8EW..#MKR.IDN.OC #:304 [309] FBB7.01.35 alpha $:304_YB8E



Message-Id: <304_YB8EW@aa5jj.bbs>
Sender: g8mny@yb8ew..#mkr.idn.oc
>From: g8mny@yb8ew..#mkr.idn.oc



From: G8MNY@YB8EW..#MKR.IDN.OC
To  : TECH@WW

Path: !VK6HGR!GB7YFS!GB7CIP!

From: G8MNY@GB7CIP.#32.GBR.EU
To  : TECH@WW

By G8MNY                                           (new Mar 06)
(8 Bit ASCII graphics use code page 437 or 850) 
I recently bought an old LAB DMM (Datron 1061a) meter. It has an unusual
feature, the ability to measure one Resistor R1 in a triangle of 3 (R2 & R3)
while all connected.

躇邊邊邊邊邊邊邊聶
糧   High��邊邊邊邊��邊邊邊聶
糧       糧    糧  R2
糧    Rx>糧   R1   ��邊邊聶
糧 DMM   糧    糧  R3  糧
糧    Low��邊邊邊邊韜邊邊邊譆  糧
糧       糧           糧
糧  Guard��邊邊邊邊邊邊邊邊邊邊邊譆
��邊邊邊邊邊邊邊譆

As the DMM has uProcessor you might expect the Guard line to be connected to
the Low then High & then disconnected, to give 3 values of Rx, then with 3
equations & 3 unknowns all the Rs can be calculated....

1/ Guard                  R1 + R2 
   connected         Rx = 邊邊邊邊邊邊邊
   to Low                 R1 x R2

2/ Guard                  R1 + R3
   connected         Rx = 邊邊邊邊邊邊邊
   to High                R1 x R3

3/ Guard                  R1 + R2 + R3
   not               Rx = 邊邊邊邊邊邊邊邊邊邊邊邊
   connected              R1 x (R2+R3)

But that is not the purpose of the Guard line in a sensitive measuring circuit.   

To measure the resistance the DMM actually uses an accurate constant current
source across Rx & measures the voltage to get the R1 value. But with R2 + R3
in circuit this will give a wrong reading as some of the accurate current will
flow through R2 + R3. But on this meter the Guard line has another current
generator with its control arranged to make the Guard the same potential as the
Low...

躇邊邊邊邊邊邊邊邊邊邊邊聶
糧       High��邊邊邊邊��邊邊邊聶
糧         / 糧    糧  R2
糧       DVM 糧   R1   ��邊邊聶
糧         \ 糧    糧  R3  糧
糧 Ref I> Low��邊邊邊邊韜邊邊邊譆  糧
糧      /糧_ /糧   No /糧\  糧
糧  躇邊邊< 糧_  糧  Voltage  糧
糧 \糧/  \糧  \糧      \糧/  糧
糧Stray I> Gu��邊邊邊邊邊邊邊邊邊邊邊譆
��邊邊邊邊邊邊邊邊邊邊邊譆

So as long as the Guard stray current generator can put the error current
through R2 the Reading of Rx = R1 & be correct.


 

Why Don't U send an interesting bul?

73 De John, G8MNY @ GB7CIP


[End of Message #643728 from G8MNY]







Read previous mail | Read next mail


 18.11.2026 01:48:58lGo back Go up