G8MNY > TECH 24.10.13 09:49l 84 Lines 3775 Bytes #999 (0) @ WW
BID : 41900_GB7CIP
Read: GUEST
Subj: Lab Measuring DMM
Path: IZ3LSV<ED1ZAC<GB7CIP
Sent: 131024/0624Z @:GB7CIP.#32.GBR.EU #:41900 [Caterham Surrey GBR] $:41900_GB
From: G8MNY@GB7CIP.#32.GBR.EU
To : TECH@WW
By G8MNY (New Aug 07)
(8 Bit ASCII graphics use code page 437 or 850, Terminal Font)
A few years ago I bought a large not working 19" LAB DMM (Datron 1061a) meter.
It is an extremely precise DMM with 6.5 digits for its time 1983, luckily it
came with all the handbooks & is quite serviceable now. (7.5 dig display)
ÚÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³ Au 10ê.1 1k 10 100 1M 10M ÚÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ³
³I+ I- êG 4w|o||o||o||o||o||o||o||o| ³ -³- | [] [] [] [] [] [] [] ³ ³
³(o)(o)(o)Ü [ ][ ][ ][ ][ ][ ][ ][ ] ³ |ø[] [] [] [] [] [] [] m A ³ ³
³ 2w 0 1 2 3 4 5 6 7 ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÙ ³
³ ³
³ Hi Lo Gu LG 8 9 . kbd ñ Cl B C Lim Lim St ³
³(o)(o)(o)Ü [ ][ ][ ] |o| [ ][ ][ ] [ ][ ][ ][ ][ ][ ] [ ][ ][ ][ ][ ] [] ³
³ RG|o||o||o| [ ] |o||o||o| |o||o||o||o||o||o| |o||o||o||o||o| [] ³
³ Z F I---fun--kê AC DC Mode Compute Pwr ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÙ
Here is part of its exacting specification...
+/- 1.999999 on most ranges with the filtering & averaging in.
e.g. with 0.19V input, you can resolve down to ñ100nV or 126dB S/N.
DC VOLTS Input Z ---- Accuracy since last Cal------
Range: 24hr, 90 day, 1 year
100mV >10Gê ñ10ppm or 0.001%, 30ppm, 45ppm
1 & 10V >10Gê ñ5ppm or 0.0005%, 20ppm, 30ppm
100V & 1kV >100Mê ñ10ppm or 0.001%, 30ppm, 45ppm
DC CURRENT ---- Accuracy since last Cal------
Range: 24hr, 90 day, 1 year
0.1-100mA ñ50ppm or 0.005%, 100ppm, 150ppm
1A ñ100ppm or 0.01%, 200ppm, 300ppm
Shunts:100mV drop across all shunts for 100% (reads to ñ199%).
RESISTANCE Test ---- Accuracy since last Cal------
Range: Current 24hr, 90 day, 1 year
10ê 10mA ñ15ppm or 0.0015%, 40ppm, 60ppm
100ê 10mA ñ10ppm or 0.001%, 30ppm, 45ppm
1kê 1mA ñ10pmm or 0,001% 30ppm. 45ppm
10kê 100uA ñ10ppm or 0.001%, 30ppm, 45ppm
100kê 10uA ñ15ppm or 0.0015% 40ppm. 60ppm
1Mê 1uA ñ30ppm or 0.003%, 100ppm, 200ppm
10Mê 100nA ñ150ppm or 0.015%, 300ppm, 500ppm
Measurement is 2 or true 4 wire system & with up to 10ê lead Resistance.
Also measures in circuit 1 of 3 Rs.
AC RMS VOLTS Input Z ---- Accuracy since last Cal------
Range 24hr, 90 day, 1 year
100mV 1Mê//150pF ñ0.04% ñ0.08% ñ0.12%
1-100V 1Mê//150pF ñ0.02% ñ0.04% ñ0.06%
100mV 1Mê//150pF ñ0.04% ñ0.08% ñ0.12%
Response flat 10Hz - 1MHz @ -2%
True RMS measurements to 7:1 crest factor.
AC RMS CURRENT ---- Accuracy since last Cal------
Range: 24hr 90 day, 1 year
0.1-1A 0.1%, 0.2%, 0.3%
Shunts: as DC
True RMS measurements to 3:1 crest factor.
Other features:
AC+DC True RMS option for Volts & Current as AC ranges.
Local or seperate screen guard system.
IEEE interface for automatic readings/logging.
Go/fail % testing.
Reading accuracy calculation.
Direct entry number keyboard.
dB calculation/readings on all ranges to ñ0.01dB.
Pre use calibrated self check.
Security key locked calibration, easy press button cal features.
See my tech bul on "Measuring 1 of 3 Rs"
Why Don't U send an interesting bul?
73 De John, G8MNY @ GB7CIP
Read previous mail | Read next mail
| |