Although you can buy a commercial probe, here is one that you can make for
either an oscilloscope or a DMM. It is designed to fit an old plastic single
piece clear pen tube. For general high voltage work there is no need for great
accuracy, just safety & kit protection.
Specifications
240V - 10kV MAX
DC to 50Hz only. Not for LOPTY AC stages!
100Mê input load.
Use 1kV rated Rs!
10:1 DIGITAL VOLTMETER PROBE
This assumes the DMM is 10Mê input.
------------- 90Mê ---------------
Wire ,/~~~³~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~³ Flexible DMM
ÄÄÄÄððððððÄ10MÄ10MÄ10M-10M-10MÄ10MÄ10MÄ10MÄ10MÄÄÄÄ===================>---Input
Probe`\___³__________________________________________³ insulated wire 10Mê
³
Earth ____________________________________________________________________³
CALIBRATION
DMM
Apply mains to DMM & note reading. Add the series probe & the reading should
have gone down to 1/10. If to high add more Rs to suit, it too low replace an R
with a low value to suit.
100:1 SCOPE PROBE 100
This assumes the scope is 1Mê input.
-------------- 99Mê ----------------
Wire ,/~~~³~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~³ Flexible Scope
ÄÄÄÄððððððÄ10MÄ10MÄ10M-10M-10MÄ10MÄ10MÄ10MÄ10MÄ9MÄÄ===================(o)Input
Probe`\___³_________________________________________³ insulated wire 1Mê
³
Earth ___________________________________________________________________³
The 9Mê is just a low value 10Mê selected on test. Or you could just use the
same DMM probe for both & recalibrate the scope to suit when used.
CALIBRATION
SCOPE
If scope has 100V/cm range, apply 240V mains at 350V peak 700V peak-peak &
measure with scope directly, it should give about 7cm p-p. Now with probe in
series adjust the scope's Y front panel gain for the same 7cm P-P display on
the 1V/cm range. Or you can adjust the R values as in DMM calibration.
MEASURING HIGH VOLTAGES
With all power off, & any HV point safely dumped to earth, connect the wire
probe. Ensure the meter/scope is safely connected to the same circuit as
needed.
With hands clear power up the HV & take readings etc.
For B/W TV/MON EHT be aware that 10kV will easily track any slightly dirty/damp
surface! Plastic sprays like "Damp Start" can be useful but be aware of non
connection to surfaces coated with it.
SAFETY
If the testgear end gets disconnected power down the HV unit & make sure any HT
is safely discharged before working on the loose testgear end of the probe.
This longitudinal resistor design should not flashover, but is extra testgear
protection is needed put a wire ended NEON tube across the DMM/SCOPE input.
That should limit the DMM or SCOPE to see only 80V (e.g. probe can only measure
up to 800V on DMM or 8kV on SCOPE).
See my Tech Buls "Scope Probes" & "Scope & DMM Calibrator"
Why Don't U send an interesting bul?
73 de John G8MNY @ GB7CIP