Subj: Lab Measuring DMM
Path: IZ3LSV<IV3YXW<I3XTY<I0OJJ<GB7CIP
Sent: 130613/0930Z @:GB7CIP.#32.GBR.EU #:26076 [Caterham Surrey GBR] $:26076_GB
From: G8MNY@GB7CIP.#32.GBR.EU
To  : TECH@WW

By G8MNY                                      (New Aug 07)
(8 Bit ASCII graphics use code page 437 or 850, Terminal Font)
A few years ago I bought a large not working 19" LAB DMM (Datron 1061a) meter.
It is an extremely precise DMM with 6.5 digits for its time 1983, luckily it
came with all the handbooks & is quite serviceable now. (7.5 dig display)

旼컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴�
�           Au 10�.1 1k 10 100 1M 10M   旼컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴커  �
쿔+ I- �G 4w|o||o||o||o||o||o||o||o|    � -�- | [] [] [] [] [] [] []       �  �
�(o)(o)(o)� [ ][ ][ ][ ][ ][ ][ ][ ]    �     |�[] [] [] [] [] [] []   m A �  �
�         2w 0  1  2  3  4  5  6  7     읕컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴켸  �
�                                                                             �
� Hi Lo Gu LG 8  9  .  kbd   � Cl                           B C Lim Lim St    �
�(o)(o)(o)� [ ][ ][ ] |o|  [ ][ ][ ]  [ ][ ][ ][ ][ ][ ]  [ ][ ][ ][ ][ ]  [] �
�         RG|o||o||o| [ ]  |o||o||o|  |o||o||o||o||o||o|  |o||o||o||o||o|  [] �
�            Z  F  I---fun--k� AC DC          Mode            Compute     Pwr �
읕컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴컴�

Here is part of its exacting specification...

+/- 1.999999 on most ranges with the filtering & averaging in.
    e.g. with 0.19V input, you can resolve down to �100nV or 126dB S/N.

DC VOLTS     Input Z   ---- Accuracy since last Cal------
Range:                     24hr,         90 day,   1 year
100mV        >10G�    �10ppm or 0.001%,   30ppm,    45ppm
1 & 10V      >10G�     �5ppm or 0.0005%,  20ppm,    30ppm
100V & 1kV   >100M�   �10ppm or 0.001%,   30ppm,    45ppm

DC CURRENT             ---- Accuracy since last Cal------
Range:                     24hr,         90 day,   1 year
0.1-100mA              �50ppm or 0.005%, 100ppm,   150ppm
1A                    �100ppm or 0.01%,  200ppm,   300ppm
Shunts:100mV drop across all shunts for 100% (reads to �199%).

RESISTANCE     Test    ---- Accuracy since last Cal------
Range:      Current         24hr,        90 day,   1 year
10�          10mA     �15ppm or 0.0015%,  40ppm,    60ppm
100�         10mA     �10ppm or 0.001%,   30ppm,    45ppm
1k�           1mA     �10pmm or 0,001%    30ppm.    45ppm
10k�         100uA    �10ppm or 0.001%,   30ppm,    45ppm
100k�        10uA     �15ppm or 0.0015%   40ppm.    60ppm
1M�           1uA     �30ppm or 0.003%,  100ppm,   200ppm
10M�        100nA    �150ppm or 0.015%,  300ppm,   500ppm

Measurement is 2 or true 4 wire system & with up to 10� lead Resistance.
Also measures in circuit 1 of 3 Rs.
 
AC RMS VOLTS  Input Z ---- Accuracy since last Cal------
Range                     24hr,      90 day,      1 year
100mV      1M�//150pF    �0.04%      �0.08%       �0.12%
1-100V     1M�//150pF    �0.02%      �0.04%       �0.06%
100mV      1M�//150pF    �0.04%      �0.08%       �0.12%

Response flat 10Hz - 1MHz @ -2%
True RMS measurements to 7:1 crest factor.

AC RMS CURRENT        ---- Accuracy since last Cal------
Range:                    24hr       90 day,      1 year
0.1-1A                    0.1%,       0.2%,        0.3%
Shunts: as DC
True RMS measurements to 3:1 crest factor.

Other features:
 AC+DC True RMS option for Volts & Current as AC ranges.
 Local or seperate screen guard system.
 IEEE interface for automatic readings/logging.
 Go/fail % testing.
 Reading accuracy calculation.
 Direct entry number keyboard.
 dB calculation/readings on all ranges to �0.01dB.
 Pre use calibrated self check.
 Security key locked calibration, easy press button cal features.


See my tech bul on "Measuring 1 of 3 Rs"

Why Don't U send an interesting bul?

73 De John, G8MNY @ GB7CIP